How Lock-In Amplifiers Beat Johnson-Nyquist Noise

This article explains the mechanics of lock-in amplification and how experimental physicists use it to recover nanovolt-scale signals buried under overwhelming thermal noise. By examining the nature of broadband Johnson-Nyquist noise, we will explore how a lock-in amplifier uses frequency modulation, phase-sensitive detection, and aggressive low-pass filtering to narrow the measurement bandwidth and extract seemingly lost data.

The Challenge of Johnson-Nyquist Noise

In experimental physics, measurements of tiny electrical signals are fundamentally limited by Johnson-Nyquist noise. This noise is generated by the random thermal agitation of charge carriers (usually electrons) inside an electrical conductor.

Because thermal noise is broadband (or "white" noise), its power is distributed equally across all frequencies. The total noise voltage you measure is proportional to the square root of your measurement bandwidth. If you use a standard wideband instrument like an oscilloscope to measure a nanovolt signal, the cumulative thermal noise across millions of Hertz will easily be thousands of times larger than your target signal, completely obscuring it.

Step 1: Modulating the Signal

To beat broadband noise, you must stop trying to measure across a broad band of frequencies. Instead of performing a static (DC) measurement, the physicist modulates the experiment at a specific, clean reference frequency (\(f_{ref}\)). This can be done by driving a circuit with an AC voltage or using an optical chopper to pulse a laser beam.

As a result, the tiny physical response you want to measure is encoded as a periodic signal oscillating at \(f_{ref}\). Meanwhile, the Johnson-Nyquist noise remains spread out randomly across the entire frequency spectrum. Furthermore, setting \(f_{ref}\) to a high frequency (e.g., above 1 kHz) moves the measurement away from low-frequency environmental noise, known as \(1/f\) noise.

Step 2: Phase-Sensitive Detection (Mixing)

The heart of the lock-in amplifier is the phase-sensitive detector (PSD), which is essentially a signal multiplier. The amplifier takes the noisy input signal from the experiment and multiplies it by an internal, noise-free reference sine wave set to the exact same frequency (\(f_{ref}\)) and phase as the experimental modulation.

According to trigonometric identities, multiplying two sine waves together produces two new signals: one at the sum of their frequencies, and one at the difference of their frequencies.

  • For the target signal: Because the target signal and the reference signal have the exact same frequency (\(f_{ref}\)), the difference between them is zero (\(f_{ref} - f_{ref} = 0\text{ Hz}\)). The multiplication converts the AC target signal into a steady DC voltage, while also creating a secondary component at twice the frequency (\(2f_{ref}\)).
  • For the thermal noise: Because Johnson-Nyquist noise exists at random frequencies and phases that do not match \(f_{ref}\), multiplying the noise by the reference wave simply shifts the noise into different AC frequencies. It does not produce a steady DC voltage.

Step 3: Low-Pass Filtering

After the multiplication stage, the mixed signal is sent through an adjustable, aggressive low-pass filter. The purpose of this filter is to average the signal over a set period (the "time constant") and block all AC signals while allowing only DC signals to pass.

This filter completely removes the \(2f_{ref}\) component generated during the mixing stage. More importantly, it strips away the broadband Johnson-Nyquist noise, which is still fluctuating as AC voltage. The only thing that survives the low-pass filter is the steady DC voltage, which is directly proportional to the amplitude of the original tiny periodic signal.

Crushing the Noise Bandwidth

Ultimately, a lock-in amplifier triumphs over thermal noise by drastically reducing the measurement bandwidth. By shifting the target signal to DC and applying a low-pass filter with a long time constant (sometimes several seconds), the lock-in amplifier creates an effective bandwidth as narrow as 0.01 Hz.

Because Johnson-Nyquist noise scales with the square root of the bandwidth, shrinking the bandwidth from megahertz down to a fraction of a Hertz eliminates almost all the thermal noise. The only noise that makes it through the filter is the microscopic sliver of thermal noise that happened to naturally occur within a fraction of a Hertz of \(f_{ref}\). This allows the lock-in amplifier to cleanly extract periodic signals that are up to a million times smaller than the surrounding noise floor.